Radionuclidic Identity
Radionuclidic Identity is confirmation that the radionuclide present in a PET drug batch is the radionuclide specified for that product.

On this page · 11 sections
- 1Radionuclidic Identity
- 2Identity, purity and chemical form answer different questions
- 3US PET requirements and dated European guidance
- 4Follow the evidence from specification to review
- 5The record relationships behind an identity result
- 6FDG-B19: complete evidence against SP-19 rev 3
- 7FDG-B19: superseded method and missing raw data
- 8A plausible identity result attached to the wrong batch
- 9Preserve evidence across integrations and recovery
- 10Controls for misleading summaries and detached approvals
- 11Write an identity conclusion the next reviewer can follow
How does Radionuclidic Identity apply to your shop floor?
Pick your industry and scale — Ask V5 rewrites the definition in your context, gives a worked example, and shows what V5 does on day one.
01Radionuclidic Identity
Radionuclidic Identity is confirmation that the radionuclide present in a PET drug batch is the radionuclide specified for that product. Radionuclide identity is an alternative name for the same attribute; radionuclidic identity test names the assessment used to establish it, not a separate quality attribute.
What radionuclidic identity means in a PET batch record
For finished-batch review, the practical question is whether the recorded identity conclusion is supported by the applicable product specification and evidence belonging to this batch. The surrounding batch record gives that conclusion its context: which product and batch were assessed, what criterion applied, and where the supporting test record can be found.
A useful review treats the identity entry as a conclusion with traceable support, rather than an isolated pass label. The reader should be able to follow its references to the applicable specification, the recorded result and the evidence on which the assessment rests. An accessible result is useful only when its relationship to the batch is clear.
This article addresses those recorded results, references, evidence relationships and review conclusions. Whether documentation is paper-based or organised in an electronic batch record, the focus is the meaning and support of the identity entry. It does not describe production or measurement procedures. The sections that follow separate the neighbouring quality attributes, then trace the evidence through practical finished-batch review cases.
02Identity, purity and chemical form answer different questions
Radionuclidic identity identifies the radionuclide present in the tested material. In a PET finished-batch record, that conclusion answers a different question from either purity result or chemical-form identification. Keeping these attributes separate helps the reviewer understand exactly what each recorded result supports.
| Record attribute | Question answered | Meaning of the recorded result | Boundary of that conclusion. |
|---|---|---|---|
| Radionuclidic identity | Which radionuclide is present? | Identifies the radionuclide against the product-specific identity criterion. | Does not itself quantify radionuclidic purity or establish chemical form. |
| Radionuclidic purity | What proportion of total radioactivity is attributable to the specified radionuclide? | Records the radionuclidic purity proportion. | Does not establish which chemical form contains that radionuclide. |
| Radiochemical identity | Which chemical form contains the radionuclide? | Identifies the chemical form containing the radionuclide. | Chemical-form evidence does not replace evidence identifying the radionuclide. |
| Radiochemical purity | What proportion of that radionuclide's radioactivity is in the specified chemical form? | Records the radiochemical purity proportion. | Does not express the proportion attributable to other radionuclides. |
An acceptable identity conclusion is therefore not a purity percentage. It records an identification assessment, not the relative contribution of the specified radionuclide to total radioactivity. Likewise, evidence that radioactivity is associated with the intended chemical form does not independently establish which radionuclide produces it. A summary that collapses these attributes into a single passing status obscures those boundaries.
The radionuclide named for the product is the intended identity; the specification's criterion is the basis for assessing the batch result. Naming the expected radionuclide in a record is not equivalent to documenting that the result meets that criterion. The distinction between label claim versus release limit is useful here: a product description and a batch acceptance criterion serve different purposes. For record review, keep the expected identity, applicable criterion and recorded assessment distinguishable rather than inferring acceptance from the product name.
03US PET requirements and dated European guidance
Binding US PET-drug regulation
For US PET drugs, 21 CFR 212.70(d) requires establishing and following procedures for the ordinary final-release pathway: “you must establish and follow procedures”. This is a procedural obligation governing final release, not a standalone definition of radionuclidic identity.
Three provisions of 21 CFR 212.70 shape the identity entry in a US PET batch record. Paragraph (a) requires specifications for each PET drug product, “including criteria for determining identity, strength, quality, purity”. Paragraph (d)(2) requires that “associated laboratory data and documentation are reviewed and they demonstrate that the PET drug product meets specifications, except for sterility”. Paragraph (d)(3) requires that “a designated qualified individual authorizes final release by dated signature.” Identity evidence therefore feeds the data review; the release authorization remains a separate signed act. Use 21 CFR 212 for the wider PET-drug context.
Dated European monograph-elaboration guidance
The EDQM Guide for the elaboration of monographs on radiopharmaceutical preparations, October 2018, is dated European monograph-elaboration guidance—not US law or a statement of current product-monograph limits. For European radiopharmaceutical monograph elaboration, EDQM guidance section 2.2.2.1 recommends an additional discriminating test when relevant radionuclides have similar characteristics.
At the batch-record level, the practical relevance is the evidence set: where the approved specification uses complementary identity characteristics, the review should address the associated results together. A summary should not conceal a missing component of that product-specific assessment. This is a suggested review approach, not an analytical procedure.
Keep three authorities separate
In the fictional FDG-B19 case, SP-19 rev 3 supplies the case’s approved identity criterion. Suggested fields—such as the method-revision reference, supporting-data link and reviewer conclusion—organise its evidence. They are a proposed record design, whether maintained on paper or in an electronic batch record. Do not present that fictional specification or field layout as regulatory text, or use it to imply a universal US radionuclidic method or acceptance number.
04Follow the evidence from specification to review
A useful PET finished-batch review follows record associations, not laboratory operations. The checkpoints below are suggested review practice, not a universally mandated sequence. They connect the identity entry in the batch record to the evidence supporting its interpretation.
- Establish the record context. Start with the batch and product identifiers, then identify the approved specification revision and the product-specific identity criterion applicable to the recorded test. A specification title alone leaves the acceptance basis unclear.
- Follow the sample association. Connect the sample identifier to that batch and to the particular identity result. This chain of identity establishes which material the result concerns; it does not itself establish which radionuclide is present.
- Connect the execution evidence. Associate the result with the method revision actually used, the analyst, the instrument record reference and the underlying data. Prefer references that resolve to the particular test record rather than a general instrument folder or laboratory workspace.
- Check historical applicability. Distinguish the executed method revision from the revision currently displayed by the document system. A present-day pointer to the current method does not establish what was used or whether that revision applied when the test was performed. Compare the execution record with the relevant approval and effective-version history.
- Close with a traceable assessment. Tie the reviewer’s identity assessment to the specific result, applicable criterion and records examined. Identify the reviewer and assessment date, and make any unresolved association explicit rather than allowing a general approval to imply that every supporting record was examined.
05The record relationships behind an identity result
A displayed radionuclidic identity conclusion is a summary, not the supporting evidence itself. A concise result entry can point to detailed records without reproducing every source record. The proposed model below makes those relationships explicit so a reviewer can follow the conclusion back to its criterion, execution and evidence.
Read the relationships in both directions: starting from the batch should lead to its result and applicable specification revision; starting from the source evidence should identify the result and batch it supports. The analyst and method revision make execution attributable, while the review reference identifies the person assessing that evidence rather than merely displaying a general approval status.
The raw_data_uri locates the underlying result evidence. The instrument_record_ref identifies the relevant instrument record. Neither substitutes for the other by its label alone: a populated reference is useful only when its destination contains the intended record and the association is clear.
Where an exception affects the identity assessment, its record links the discrepancy, affected evidence and reviewer's assessment. This gives the reviewer a place to explain its significance instead of leaving an unexplained flag beside the result.
These field names and entity boundaries are design suggestions, not legally prescribed terminology. In an electronic batch record, equivalent controlled fields or references can express the same relationships; the practical test is whether the reviewer can follow them without guessing.
Batch
Anchors product and sample associations.
- batch_id
- product_id
- sample_id
Specification
Defines the applicable product-specific identity criterion.
- specification_id
- specification_revision
- specified_radionuclide
- identity_criterion
- approved_method_ref
Identity result
Connects batch, specification and evidence.
- identity_result_id
- batch_id
- sample_id
- specification_ref
- identity_result
- evidence_ref
Source evidence
Identifies execution and supporting records.
- evidence_ref
- method_id
- method_revision
- analyst_id
- instrument_record_ref
- raw_data_uri
- recorded_at
Review
Records assessment and separate disposition reference.
- review_id
- identity_result_id
- reviewer_id
- reviewed_at
- identity_assessment
- release_record_ref
Exception
Links discrepancies to their documented assessment.
- exception_id
- identity_result_id
- issue_type
- assessment_ref
- resolution_status
06FDG-B19: complete evidence against SP-19 rev 3
FDG-B19 is a fictional PET finished batch, and SP-19 rev 3 is its fictional product-specific specification. Assume that this specification was approved and applicable when testing occurred, that the tested sample belongs to FDG-B19, and that the referenced approved method revision is applicable under that specification. SP-19 rev 3 defines a radionuclidic identity criterion using a combination of identifying characteristics. This example concerns the recorded evidence supporting that criterion, not an analytical procedure.
A shared annihilation signal alone cannot distinguish positron emitters. In fictional SP-19 rev 3, the identity conclusion therefore depends on the specified discriminating evidence, rather than that shared signal alone. A recorded half-life characteristic contributes to the specified evidence set; it is considered alongside the other identifying characteristics referenced by the specification. The reviewer follows those evidence references rather than treating a generic signal description as sufficient identification.
| Record entry | Completed fictional Branch A entry |
|---|---|
| Identity result | “Specified radionuclide identified; SP-19 rev 3 combined identity criterion met” — entry ID-B19-01 |
| Specification | SP-19 rev 3, effective before the FDG-B19 test date, attached to ID-B19-01 |
| Approved method revision | Method QC-RID-04 rev 2, listed as current under SP-19 rev 3 |
| Analyst | Analyst A. Mensah, entry time-stamped and attributed to ID-B19-01 |
| Instrument record reference | Instrument log entry INS-B19-07 for the system used on the FDG-B19 sample |
| Raw data link | Analytical record AR-B19-01 (sample FDG-B19-S1), opened by the reviewer |
| Reviewer | QC reviewer J. Patel, identity assessment completed with reference to ID-B19-01, AR-B19-01 and INS-B19-07 |
| Separate release-record reference | Release record REL-B19, linked for the designated individual’s separate decision; no outcome stated here |
The entries form a connected assessment: the result addresses the applicable criterion, the method reference agrees with the specification, and the accessible analytical record supports the reported conclusion for the assigned sample. Any reported measurement and its unit remain attributable to that analytical record. No numerical result, acceptance range or arithmetic is introduced into this qualitative example.
Branch A concludes with the reviewer recording identity evidence acceptable against SP-19 rev 3, with references to the evidence reviewed. Release record REL-B19 is referenced, not decided, by this entry.
07FDG-B19: superseded method and missing raw data
In this fictional Branch B, PET batch FDG-B19 is assessed against product-specific specification SP-19 rev 3. Its identity result entry references a superseded method revision, lacks a working raw data link, or contains both discrepancies. No resolved explanation establishes which evidence applies. The initial identity-review status is incomplete, even if the result entry reports that the identity criterion was met.
The superseded reference leaves two materially different possibilities: the analyst entered the wrong revision identifier, or testing actually used a revision that was not applicable. The proposed review approach records the discrepancy and examines the execution history, including the analyst’s record, method association and instrument record reference. Silently replacing the revision would hide the question rather than answer it. Any correction retains the original entry and explains the evidence supporting the change.
Similarly, a missing link does not demonstrate that the underlying data are lost. It establishes that the reviewer cannot currently follow that connection. The assessment distinguishes an inaccessible reference from absent, incomplete or incorrectly associated data. Identity review remains incomplete while the nature and extent of the gap are assessed; technical recovery belongs to the integrations discussion.
For this case, the exception record identifies the issue, affected identity entry and assessment owner. It records the evidence examined, unresolved questions, resolution when established, and subsequent reviewer assessment. A repaired link or corrected identifier is not itself the conclusion: the reviewer reassesses whether the resulting evidence supports the identity entry against SP-19 rev 3.
An incomplete record alone does not establish a failing analytical result. Where the result fails the applicable criterion, the separate out-of-specification investigation pathway applies. Neither documentation gap implies automatic rejection, acceptance or conditional release. Branch B ends with a documented investigation path and an incomplete review status until the record supports a different assessment.
- Bind batch and specificationAssociate FDG-B19 with applicable SP-19 rev 3.
- Evidence complete?Yes: compare criterion. No: investigate the documentation gap.
- Compare criterionMeets: accept identity evidence. Fails: assess the failing result.
- Investigate documentation gapAssess revision or linkage discrepancies; return for evidence review.
- Assess failing resultFollow the documented OOS pathway; do not infer disposition.
- Accept identity evidenceReviewer records the supported assessment against the applicable specification.
- Reference disposition separatelyLink the release decision without converting identity acceptance into approval.
08A plausible identity result attached to the wrong batch
In this additional fictional case, FDG-B19 has an identity attachment that opens successfully and contains an apparently acceptable conclusion. However, the attachment’s sample identifier maps to a different batch. Accessibility and a plausible result do not establish applicability to FDG-B19. This discrepancy does not establish that FDG-B19 contains the wrong radionuclide; it establishes a problem with the evidence association.
Using the evidence relationships described above, the reviewer compares the attachment’s identifier with the sample-to-batch association in the batch record. The reviewer records both identifiers, identifies the attachment involved and documents where the relationship breaks. The identity-evidence acceptance conclusion for FDG-B19 remains withheld while the applicable evidence is established. An acceptable conclusion for another batch cannot fill that gap.
Resolution starts with establishing which record actually belongs to FDG-B19, rather than relabelling the convenient attachment. If the assessment confirms an attachment-selection error, correcting that association is distinct from changing an analytical result. The original result remains associated with its supported batch. Any proposed correction to an identifier within the analytical record needs its own documented basis; it is not justified merely because the destination batch needs evidence.
A useful resolution record connects the original attachment association, the discrepancy assessment, the reason for correction and the applicable replacement evidence. It also identifies who reviewed the corrected relationship. Once that relationship is supported, the reviewer can assess FDG-B19’s identity evidence against its specification. The separate release decision record can reference this resolved assessment without treating attachment repair itself as an identity result or batch-release approval.
09Preserve evidence across integrations and recovery
As a system-neutral design practice, connect the electronic batch record to the laboratory result store, controlled specification and method records, and review or disposition records through identifiable record references. These interfaces should let a reviewer follow the identity entry to its supporting evidence without assuming that matching display labels establish the relationship.
Preserve stable identifiers, the specification revision applied, the method revision actually executed and the original result status when transferring records. A displayed summary should retain an explicit relationship to its underlying result and supporting data. If a receiving system translates status labels, document their meaning so that an imported entry does not appear reviewed merely because transfer succeeded.
For a broken reference, locate the attributable source record and establish its relationship to the affected identity result before repairing the association. Document what changed, who made the repair and its basis, then reassess access from the reviewer’s normal record view. Keep both the original association and the correction intelligible to later reviewers. An unrelated replacement attachment is not recovery, even if its contents look plausible.
Treat an inaccessible or obsolete file representation as an access problem requiring assessment. Consider whether the available representation preserves the result’s attribution, content and context. No universal file format resolves every case, and successful conversion alone does not establish that evidence remains interpretable. Record unresolved limitations rather than presenting a readable substitute as equivalent without assessment.
10Controls for misleading summaries and detached approvals
A readable summary can still leave the identity assessment ambiguous. The following practical controls help a reviewer understand what was assessed, against which criterion, and on what evidence. They are implementation suggestions, not prescribed record formats.
- Replace generic status with a scoped assessment. An acceptable status alone does not identify the quality attribute assessed. Pair it with the explicit radionuclidic identity test name and applicable specification criterion reference, keeping both visible wherever the result is summarised.
- Attach approval to its evidence. A reviewer signature displayed separately from the result can leave its scope unclear. Associate the assessment with the specific result and evidence set reviewed, so another reviewer can distinguish approval of that assessment from an unrelated record approval.
- Make checklist completion batch-specific. A copied-forward checklist can look complete without showing a fresh assessment. Use evidence references belonging to the batch and a meaningful conclusion describing what those references support, rather than relying on inherited ticks or generic completion text.
For chemical-form confusion, use the earlier comparison. For method discrepancies or missing evidence, follow the earlier exception cases; use the OOS discussion for failing results. These controls make the assessment traceable without substituting summary presentation for evidence review.
11Write an identity conclusion the next reviewer can follow
In fictional Branch A, the reviewer assessed the linked identity evidence for FDG-B19 against SP-19 rev 3 and recorded it as acceptable. The conclusion references the identity result entry, approved method revision, analyst attribution, instrument record reference and raw data link already associated with FDG-B19. It retains the reviewer's identity and assessment scope without inventing additional identifiers or analytical values.
Frequently asked questions
Q.Can the same source record support separate identity and [radionuclidic purity](/glossary/radionuclidic-purity) entries?+
Yes, where the approved product-specific methods and specification support both uses. As a record-review practice, each entry should identify its own criterion, result and assessment, even when both reference the same underlying data. Identity addresses whether the specified radionuclide is present; radionuclidic purity addresses the proportion attributable to that radionuclide. A shared attachment does not make those conclusions interchangeable. Cross-reference the relevant portions of the source record so the reviewer can follow each assessment independently.
Q.Which method revision belongs in the identity record when a new revision takes effect after testing?+
Record the method revision actually used for testing. A later effective revision should not silently replace that historical reference. As a review practice, assess applicability against the controlled implementation instructions, including any transition provisions affecting work already completed. If the change affects interpretation of the identity result, document that assessment and its disposition separately. The batch record should distinguish the method used from any later revision considered during review, rather than presenting them as equivalent.
Q.Does the raw data file need to be copied into the batch record, or can the record reference it?+
A controlled reference can be a practical alternative to duplicating a raw data file, provided the reviewer can retrieve the correct retained record and understand its batch and test association. The reference should resolve to identifiable evidence, not merely a folder or changeable search result. In an electronic batch record, useful reference details include the source record identifier, relevant version and location. Copying a file alone does not establish that it is complete or authoritative.
Q.How should a reviewer handle identity evidence that exists but cannot be read in the review system?+
Treat unreadable evidence as an unresolved review-access problem, not as an acceptable identity assessment. Record the affected source identifier, the access or rendering problem, and the review work that remains incomplete. An alternative controlled viewing route may allow assessment, but document what was actually examined and how it connects to the batch. Under the site's documentation workflow, distinguish this problem from an out-of-specification result: inability to read evidence does not itself demonstrate that the identity criterion failed.
Q.What should be recorded when an identity review is handed over to another reviewer?+
A useful handover identifies the outgoing and incoming reviewers, the evidence already examined, outstanding questions, and the current assessment status. Include the handover time and references to any unresolved documentation issue or investigation. Avoid a generic review-in-progress note that leaves the next reviewer guessing which checks were completed. The electronic release record can reference this handover without implying completion. The incoming reviewer should make their own contribution attributable rather than signing as though they performed the earlier review.
Q.How should a corrected identity entry remain connected to the original entry and assessment?+
As a traceability practice, retain the original entry and connect the correction through a recorded reason, author and time. Identify whether the correction changes only documentation or changes the interpretation of the identity evidence. If an earlier assessment relied on the corrected content, record the reassessment and its relationship to that earlier conclusion. An audit trail review workflow helps structure this examination; the correction should not leave two apparently current identity conclusions with no explanation of their relationship.
Q.Does identity evidence have to be reviewed before sterility results are available?+
For the ordinary final-release pathway in 21 CFR 212.70(d), paragraph (d)(2) requires that associated laboratory data and documentation are reviewed and demonstrate that the PET drug product meets specifications, “except for sterility”. The exception named is sterility; identity is part of the specification evidence reviewed under that paragraph. Conditional release under 212.70(f) is a separate pathway, covered in conditional release.
Q.How should an identity result from a contract laboratory appear in the PET batch record?+
Record the laboratory, its report identifier and version, and the sample identifier that ties the report to the batch, alongside the product-specific criterion it was assessed against. The firm's own reviewer still records whether that report supports the identity entry. An outsourced test changes where the source evidence is held, not the need for a traceable, attributable assessment in the batch record.
Primary sources
Want to see how Radionuclidic Identity could fit into your own records and workflows? Explore the related V5 pages or talk to our team about what applies to your operation.