V5 Ultimate
Ultimate
PricingResourcesCompany
English (original)العربيةDeutschΕλληνικάEspañolFrançaisItalianoעברית日本語한국어NederlandsPolskiPortuguêsSvenska
Talk to us Log in Start free trial
Food & Beverage
  • Bakery & Confectionery
  • Food Processing
  • Ingredients & Dry Mixes
  • Sausage & Meat Products
  • Produce Packing
  • Pet Food & Pet Supplements
  • Dairy & Chilled Foods
  • Beverages & Liquid Manufacturing
  • Infant Formula
Pharma & Life Sciences
  • Radiopharmaceuticals
  • Pharmaceutical
  • Medical Devices
  • Veterinary Pharmaceuticals
  • Blood & Tissue
  • Diagnostics & IVD Kits
  • Reference Materials & Standards
  • OTC Topical Creams & Gels
Supplements & Personal Care
  • Dietary Supplements
  • Cosmetics & Personal Care
  • Cannabis & Hemp Manufacturing
  • Fragrances & Perfumery
  • Soaps & Personal Hygiene
  • Botanical Extracts & Essential Oils
Chemicals & Materials
  • Plastics & Resins
  • Agricultural Chemicals
  • Specialty & Industrial Chemicals
  • Printing Inks, Pigments & Construction Chemicals
  • Lubricants & Metalworking Fluids
  • Coatings, Adhesives & Sealants
Other Manufacturing
  • Consumer Products
  • Electronics & Electrical Assembly
  • Textiles & Nonwoven
  • Contract Manufacturing & Co-Packing
  • Packaging & Food-Contact Materials
  • Regulated Warehousing & Distribution
Explore all industries
  • Warehouse & TraceabilityReceive, store, pick and trace
  • Production & PlanningPlan jobs and guide production
  • Quality & LaboratoryTest, investigate and review
  • Documents & TrainingControl procedures and qualifications
  • Equipment & PerformanceMaintain assets and measure losses
  • Orders & IntegrationsConnect customers, suppliers and systems
  • Outputs & InsightsReports, KPIs and label design
Why V5? Product & Features
Why V5?Product & Features
  • Warehouse & TraceabilityReceive, store, pick and trace
  • Production & PlanningPlan jobs and guide production
  • Quality & LaboratoryTest, investigate and review
  • Documents & TrainingControl procedures and qualifications
  • Equipment & PerformanceMaintain assets and measure losses
  • Orders & IntegrationsConnect customers, suppliers and systems
  • Outputs & InsightsReports, KPIs and label design
Explore all industries
    • Bakery & Confectionery
    • Food Processing
    • Ingredients & Dry Mixes
    • Sausage & Meat Products
    • Produce Packing
    • Pet Food & Pet Supplements
    • Dairy & Chilled Foods
    • Beverages & Liquid Manufacturing
    • Infant Formula
    • Radiopharmaceuticals
    • Pharmaceutical
    • Medical Devices
    • Veterinary Pharmaceuticals
    • Blood & Tissue
    • Diagnostics & IVD Kits
    • Reference Materials & Standards
    • OTC Topical Creams & Gels
    • Dietary Supplements
    • Cosmetics & Personal Care
    • Cannabis & Hemp Manufacturing
    • Fragrances & Perfumery
    • Soaps & Personal Hygiene
    • Botanical Extracts & Essential Oils
    • Plastics & Resins
    • Agricultural Chemicals
    • Specialty & Industrial Chemicals
    • Printing Inks, Pigments & Construction Chemicals
    • Lubricants & Metalworking Fluids
    • Coatings, Adhesives & Sealants
    • Consumer Products
    • Electronics & Electrical Assembly
    • Textiles & Nonwoven
    • Contract Manufacturing & Co-Packing
    • Packaging & Food-Contact Materials
    • Regulated Warehousing & Distribution
PricingResourcesCompany
Start free trial Log in Talk to us
HomeGlossaryDevice Relabelling
Compliance · The complete guide

Device Relabelling

In short

A practical reference for controlling translated device information, affected stock and documented human decisions before destination-market distribution.

4,247 words · ~20 min read
Device relabelling under EU MDR Article 16: a distributor operator in hair cover, mask, glasses and gloves applies a translated label to a device carton while a QA reviewer checks it against the manufacturer's original on a tablet.
Relabelling evidence ties each translated label to its manufacturer source, the affected stock and the destination market.
On this page
  1. 01Device Relabelling
  2. 02Screen manufacturer responsibility before choosing the Article 16 pathway
  3. 03Compare relabelling, artwork approval, reconciliation and repackaging
  4. 04Build the QMS record around source accuracy and affected stock
  5. 05Assemble notification, sample and notified-body certificate evidence
  6. 06Run the relabelling job from intake to human disposition
  7. 07Connect device scope, revisions, stock and market evidence
  8. 08Work through a market release, a revision hold and a packaging stop
  9. 09Recover exceptions without crossing the original-condition boundary
  10. 10Keep the EU medical-device rules separate from adjacent frameworks
  11. 11Review one complete file before authorising distribution
On this page · 11 sections
  1. 1Device Relabelling
  2. 2Screen manufacturer responsibility before choosing the Article 16 pathway
  3. 3Compare relabelling, artwork approval, reconciliation and repackaging
  4. 4Build the QMS record around source accuracy and affected stock
  5. 5Assemble notification, sample and notified-body certificate evidence
  6. 6Run the relabelling job from intake to human disposition
  7. 7Connect device scope, revisions, stock and market evidence
  8. 8Work through a market release, a revision hold and a packaging stop
  9. 9Recover exceptions without crossing the original-condition boundary
  10. 10Keep the EU medical-device rules separate from adjacent frameworks
  11. 11Review one complete file before authorising distribution
AI · Explain it for MY operation

How does Device Relabelling apply to your shop floor?

Pick your industry and scale — Ask V5 rewrites the definition in your context, gives a worked example, and shows what V5 does on day one.

Your scale

01Device Relabelling

Device Relabelling is the provision or replacement of medical device labelling or instructions for use, including translations of manufacturer-supplied information for a destination market. Device relabeling is the American-spelling alias; this article uses relabelling.

Operationally, the work may concern a device label, accompanying instructions for use (IFU), or both. The destination Member State identifies the intended EU market, but a shared destination does not make every information or packaging change the same activity. Describe what is being changed before deciding which regulatory pathway applies.

The practical record question is: which manufacturer source information, local revision, affected devices and market decision belong together? A useful file connects the source label or IFU to the translated or replacement version, identifies the affected stock, and records the human decision authorising its distribution. Device identifiers, including the UDI where applicable, can help connect that information to the intended product. Keep the original manufacturer’s identity separate from the identity of the organisation performing the work; document both roles rather than treating them as interchangeable.

The focus here is EU MDR medical-device relabelling by distributors and importers, particularly the Article 16 pathway. The discussion distinguishes legal obligations from guidance and proposed record controls; its conclusions are not a general rule for other regulated products or jurisdictions.

02Screen manufacturer responsibility before choosing the Article 16 pathway

Start with who will carry manufacturer obligations, because that decides whether the Article 16 relabelling route is even available. Under Article 16(1), a distributor, importer or other person assumes the obligations incumbent on manufacturers if it does any of three things, set out below with the exceptions the same paragraph provides.

Article 16(1) triggerWhat it coversException or limit in the textRelabelling screen question
(a) Own nameMakes available a device under its name, registered trade name or registered trade markNot where the distributor or importer has an agreement with the manufacturer whereby the manufacturer is identified as such on the label and is responsible for the manufacturer requirementsWill our name appear as if we were the manufacturer, and if so, is there such an agreement?
(b) Intended purposeChanges the intended purpose of a device already placed on the market or put into serviceNone stated for relabelling; translated text must not widen or narrow the indicationsDoes any translated or added text change what the device is for?
(c) ModificationModifies a device already placed on the market or put into service so that compliance with the applicable requirements may be affectedArticle 16(2) lists what is not such a modification (next paragraphs)Is the change confined to 16(2)(a) information or 16(2)(b) outer packaging?
Individual patientAssembling or adapting for an individual patient a device already on the marketParagraph 1 does not apply to a person who does this without changing its intended purposeRarely relevant to market relabelling; record if claimed

Article 16(2) then narrows trigger (c). Point (a) covers provision, including translation, of the information supplied by the manufacturer, in accordance with Section 23 of Annex I, relating to a device already placed on the market, and of further information which is necessary in order to market the device in the relevant Member State. The Annex I reference matters: the translated or added information must still meet the manufacturer-information requirements of Section 23, so the screen should record that the local text is checked against that section rather than against the distributor's own preferences.

Point (b) covers changes to the outer packaging of a device already placed on the market, including pack size, if the repackaging is necessary to market the device in the relevant Member State and is carried out so that the original condition of the device cannot be affected. Anything outside (a) and (b) leaves trigger (c) open and therefore the possibility that the performer becomes the manufacturer, with every manufacturer obligation, rather than following the Article 16(3)–(4) route. The manufacturer of the device remains its manufacturer in the Article 16(2) route; the distributor or importer adds its own obligations, it does not take over the manufacturer's.

Work done for the manufacturer

One arrangement sits outside this route altogether. Non-legally-binding MDCG 2021-26 notes that Article 16(2), (3) and (4) do not apply to operators subcontracted by the manufacturer, who carry out relabelling or repackaging on behalf of and under the control of the manufacturer, even if they also qualify as importers or distributors. That is different from the 16(1)(a) own-name agreement exception: in the subcontract case the activity is the manufacturer's own operation under its quality system; in the agreement case the distributor is marketing under its own name with the manufacturer still identified on the label. Record which operating arrangement applies, with the contract reference, before choosing the evidence set described in the rest of this article.

As a practical control, create a responsibility-screen record before commissioning translation or packaging work. Keep the legal-status assessment separate from the production instruction, using these fields:

  • Activity description: what information or packaging changes, and what remains unchanged.
  • Market rationale: destination Member State and the reason the proposed activity is needed.
  • Manufacturer identity: the organisation identified as manufacturer for the device.
  • Performer identity: the organisation translating, printing, applying labels or changing packaging.
  • Reviewer conclusion: proposed pathway, reasons, unresolved questions and the responsible human reviewer.

Where pack configuration changes, link the relevant package device identifier record. Do not infer manufacturer status merely from which organisation printed or applied a label; record any unresolved status question before proceeding.

03Compare relabelling, artwork approval, reconciliation and repackaging

These activities commonly share records, but answer different questions. Treating a completed proof, translation or label count as the whole relabelling decision leaves gaps between approved information, physical stock and the intended destination market.

ActivityDecision addressedUseful evidenceWhat it does not settle
Device relabellingWhich information change applies to which devices and market?Source revision, affected stock, destination and dispositionIndividual checks without their supporting records
Artwork approvalIs content and layout acceptable?Approved proof, comments and revision identityPhysical label accountability or market readiness
Label reconciliationAre issued labels accounted for?Issued, applied, returned and destroyed quantitiesWhether label content is correct
Translation reviewDoes the target text convey the source accurately?Paired texts, terminology decisions and reviewer approvalWhether the source revision remains current
Outer-packaging changeIs the proposed packaging configuration suitable?Configuration specification, packing instructions and inspection resultsWhether the destination-market decision is complete

Use artwork proof approval for content and layout review. For label reconciliation, connect the quantity record to a batch label specimen: balanced quantities alone cannot identify the version actually applied. Translation review likewise benefits from an explicit source-revision reference rather than a language approval detached from its source.

Treat the US term initial importer as a separate jurisdiction-specific terminology question, not another name for the EU relabelling performer.

04Build the QMS record around source accuracy and affected stock

Article 16(3) begins with an obligation that appears on the product, not in the file. A distributor or importer carrying out 16(2)(a) or (b) activities shall indicate on the device or, where that is impracticable, on its packaging or in a document accompanying the device, the activity carried out together with its name, registered trade name or registered trade mark, registered place of business and contact address, so that its location can be established (Article 16(3)). The device itself is the default; only where marking it is impracticable may the statement go on the packaging or in an accompanying document, which the text offers as alternatives. Specimen review should check this statement as well as the translated content, and record why marking the device was impracticable when another placement is used.

The second subparagraph of Article 16(3) requires a quality management system with procedures ensuring that translation of information is accurate and up to date.

The same subparagraph requires procedures ensuring the activities are performed under conditions that preserve the original condition of the device and that repackaged packaging is not defective, of poor quality or untidy, and procedures ensuring the distributor or importer is informed of manufacturer corrective action relating to the device to respond to safety issues or bring it into conformity.

Translate procedural outcomes into a usable record

The following fields are a proposed implementation design, not a prescribed Article 16 schema. They connect document control and revision control to the stock on which a particular translation is used. Keep content approval separate from permission to use that content on identified devices.

Proposed fieldsReview question
Manufacturer source revision; local revisionWhich source supports this translation, and which local version implements it?
Language; destination marketWhere and in which language is this version intended for use?
Device identifier; affected lot or serial scopeWhich physical devices are associated with the approved information?
Translation reviewer; approval rationaleWho assessed the translation, and why was it accepted?
Effective-use statusIs this version pending review, approved for defined use, or no longer authorised for new use?

For traceability, identify the packaging level when recording a package device identifier; a device-family description alone may leave the affected presentation unclear. Where an electronic device history record forms part of the record architecture, link the relabelling record to the relevant stock history rather than copying uncontrolled revision descriptions between files.

Keep comparison, competence and approval evidence distinct

A source-to-translation comparison answers whether the local text preserves the source meaning and covers the intended content. A reviewer competence record answers why that person was suitable to assess the language and device terminology. Controlled approval evidence answers who accepted the identified revision, when, for what scope and on what rationale. None substitutes for the others.

Use attributable recording to connect review comments and decisions to responsible people, and assign ownership of incoming manufacturer corrective-action notices so they reach the affected content revisions and stock.

05Assemble notification, sample and notified-body certificate evidence

For EU MDR medical devices undergoing Article 16(2)(a) or (b) activities, Article 16(4) requires distributors or importers to inform the manufacturer and the destination Member State’s competent authority of their intention at least 28 days before making the relabelled or repackaged device available on the market.

For EU MDR medical devices in this pathway, Article 16(4) also requires the distributor or importer, upon request, to provide the manufacturer and competent authority with a sample or mock-up of the relabelled or repackaged device, including any translated label and instructions for use. This sample provision is request-dependent; the advance-information obligation is not.

For EU MDR medical devices in this pathway, Article 16(4) requires submission of a certificate to the competent authority within the same period of 28 days. The certificate must be issued by a notified body designated for the relevant device types and attest that the distributor’s or importer’s QMS complies with Article 16(3).

Keep transmission evidence separate from release

A practical evidence register can use the following fields. This is a proposed record structure, not a prescribed regulatory form. Keep separate entries for each recipient so that a message to one party cannot stand in for evidence concerning the other.

Register fieldEvidence to connect
RecipientManufacturer or competent authority; contact and destination Member State.
Device and market scopeDevice identifiers, affected stock reference, activity, language and intended market.
Transmitted versionExact notification, label and IFU revisions included in the transmission.
Timing and deliveryTransmission date, intended availability date and supporting delivery evidence.
Sample request statusRequest received or none recorded; requested material, response version and dispatch evidence.
Certificate referenceCertificate identifier and issuance record, separately linked to authority-submission date and delivery evidence.

Certificate issuance and certificate submission answer different review questions: what document exists, and what was transmitted to the authority. Keep both records rather than treating the certificate’s issue date as evidence of submission. Where an electronic device history record holds the relabelling execution record, a stable reference can connect it to this register without duplicating attachments.

The human reviewer checks recipient coverage, version consistency, delivery evidence, request handling and the timing calculation; the release decision itself is described once, in the operational section.

06Run the relabelling job from intake to human disposition

The job runs through the stages in the diagram below. Ownership is a local choice; what matters is that each handoff names the same job and revision, and that a changed source or stock scope goes back to the earlier stage rather than being silently substituted.

This is the principal home for the release rule used throughout the article: stock status stays unchanged until an authorised person records the quality-release decision, and neither a sent notification, the elapsed 28 days nor a certificate attachment is itself that decision. Execution checks against approved content and work instructions are captured in the batch record before that review.

From destination request to human disposition
  1. 1
    Screen scopeRecord the activity, destination and responsibility assessment.
  2. 2
    Baseline informationIdentify manufacturer documents and their applicable revisions.
  3. 3
    Review translationConnect local wording with its source and reviewer.
  4. 4
    Control executionAssociate affected stock with the job and checks.
  5. 5
    Assemble market evidenceCollect notifications, certificate references and requested samples.
  6. 6
    Assess readinessReview timing, evidence consistency and unresolved exceptions.
  7. 7
    Record dispositionAn authorised person documents the decision and scope.
Each handoff connects the proposed activity with controlled information, affected stock and review evidence.

07Connect device scope, revisions, stock and market evidence

A stable device-scope identifier covering device variant, destination market and language is the anchor; translation revisions, jobs, stock and market evidence attach to it, so that a notification supports a specific scope rather than acting as organisation-wide permission.

Define handoff contracts around master data and lot genealogy using data integrity by design. The following is a proposed allocation of identifiers, revision references and status ownership.

HandoffReferences transferredStatus owner and exception route
Controlled documents → translation filesScope ID, source document ID/revision, language, translation ID/revisionDocument control owns source status; mismatches return for assessment.
Translation files → job executionJob ID, approved translation revision, label/IFU revisionJob supervisor owns execution status; duplicate references go to reconciliation.
Inventory → job executionDevice ID, lot/serial associations, stock status, inventory transaction IDInventory owner resolves missing associations before affected stock proceeds.
Job execution → quality fileJob ID, execution-record revision, stock links, market-evidence IDs/versionsQA owns disposition; incomplete transfers enter an exception queue.

For controlled record handoffs, design acceptance checks to reject incomplete imports without silently dropping fields. Retain the rejection reason and original payload reference. Distinguish a retry from a genuinely new job before accepting duplicate job references. Route missing lot associations to the inventory owner and mismatched document revisions to document control; neither should be repaired by guessing. Record the correction and obtain acceptance before resuming the affected handoff.

Preserve reviewer identity, decision time and the evidence version used, and link them to the electronic release record.

A connected relabelling evidence model

Device scope

Identifies the device, stock and intended destination.

  • scope_id
  • device_id
  • lot_or_serial_scope
  • destination_member_state

Translation revision

Connects local content with its manufacturer source.

  • source_revision_id
  • local_revision_id
  • language
  • reviewer_id

Relabelling job

Associates execution with affected stock and checks.

  • job_id
  • scope_id
  • local_revision_id
  • inspection_record_id

Market evidence

Records communications and certificate associations.

  • evidence_id
  • recipient
  • submitted_at
  • certificate_id

Release decision

Preserves the human disposition and reviewed scope.

  • decision_id
  • scope_id
  • evidence_set_id
  • approver_id
  • decided_at
Proposed fields connect information and stock; they are not a statutory database schema.

08Work through a market release, a revision hold and a packaging stop

These fictional EU medical-device cases use anonymised identifiers to illustrate a proposed human review workflow. They are not customer results. Apply the earlier Article 16(4) advance-information explanation when reviewing the timing evidence; the examples below focus on how the decision file fits together.

Fictional case REL-ALPHA: planned market availability

Assume the activity has been assessed as within the chosen pathway. DEV-ALPHA is non-sterile; its manufacturer information, approved translation IFU-LOCAL and affected stock LOT-ALPHA are identified. Both notifications—to the manufacturer and competent authority—and certificate submission occur on the same recorded calendar date, D.

Calendar checkpointFictional recordReview meaning
Notification and submissionDBoth notification records and the certificate-submission record carry this established date.
Proposed availabilityD + 28 calendar daysPlanned availability date, not an approval event.
Elapsed interval(D + 28 calendar days) − D = 28 daysCalendar calculation for this example.

The example assumes the relevant notification and submission dates are established. It does not determine legal service or time-computation conventions. Retain the underlying correspondence rather than treating a date entered in a planning field as sufficient timing evidence.

The reviewer assesses that timing evidence alongside the IFU-LOCAL revision baseline, LOT-ALPHA stock scope, execution checks in the batch execution history and the certificate reference, and records a disposition naming the stock and planned date. In this case the device label also carries the distributor's Article 16(3) activity statement, so that element is checked on the specimen, not assumed from the artwork file.

Fictional case REL-BETA: source revision before disposition

Assume a new manufacturer source revision arrives after translation approval but before planned disposition. Place REL-BETA on hold. Compare the previous source, revised source and approved local text, recording which passages changed and whether their meaning affects the destination-language information.

Assess stock separately: identify unprocessed devices, completed relabelled stock and unused printed material, using perpetual inventory records to locate affected quantities. For this fictional case, assume the revision changes a warning reflected incorrectly in the local text. The updated human decision retains the hold pending corrected translation, controlled replacement and renewed review of market evidence. Its rationale identifies the warning change and affected stock. A formatting-only revision could lead to a different documented decision; revision arrival alone does not establish the consequence.

Fictional case REL-GAMMA: uncertain packaging boundary

Assume the proposed job involves a sterile device, but the team cannot establish which packaging maintains sterility. Stop REL-GAMMA before opening or replacing packaging. Record the uncertainty, preserve the stock’s existing condition and request a packaging and responsibility assessment. The human decision remains a stop pending resolution, not permission inferred from packaging appearance. Use the boundary analysis in the next section to frame that assessment.

09Recover exceptions without crossing the original-condition boundary

For EU MDR medical devices placed on the market sterile, non-legally-binding MDCG 2021-23, footnote 3, reproducing Article 16(2)(b), explains the presumption that original condition is adversely affected if repackaging opens, damages or otherwise negatively affects packaging necessary to maintain sterility.

Failure modeProposed focused recovery
Stale source informationCompare revisions and review the impact on translated content and affected stock.
Untraceable stockSegregate stock; reconstruct device, lot and job links before disposition.
Incorrect local contentAssess which devices, language versions and destinations share the error.
Uncertain packaging conditionStop work; obtain a documented packaging assessment before considering resumption.

Use quarantine and controlled stock recovery to separate suspect units, with their quarantine status visible in inventory. Record the deviation with affected scope, containment, findings, revised instructions and disposition, and keep the original job linked to its recovery record.

Route recurring causes into corrective action and instruction changes through change control; unresolved patterns can inform management review. A named human reviewer should document why the checks support restarting the affected activity. These are risk-based implementation suggestions, not a universal sampling plan, retention period or acceptance threshold.

10Keep the EU medical-device rules separate from adjacent frameworks

For EU medical devices, the binding framework discussed here is Regulation (EU) 2017/745, the EU MDR. Keep three functions separate when reviewing a relabelling file:

  • Binding provisions establish the applicable legal obligations; consult the EU MDR text for the legal decision.
  • For EU device relabelling, MDCG 2021-26 and MDCG 2021-23 are non-legally-binding guidance, not substitutes for the legislation.
  • The proposed record designs in this article are operational suggestions: they organise evidence for human review, rather than prescribe a legally mandated form or confer compliance status.

For in vitro diagnostic devices, check the applicable IVDR framework rather than transferring this article’s MDR conclusions. For US device-import activities, assess the initial importer role under the applicable US framework; do not treat the EU distributor/importer pathway as its equivalent.

As a practical review convention, identify the market, product framework and operator role before selecting a procedure or reusing a record template.

11Review one complete file before authorising distribution

Use this practical coherence check alongside the scope and evidence explanations above, not as a substitute for them. Select a device or lot and follow its source information, local revision, relabelling job, market evidence and disposition.

  • Is the activity assessment recorded and linked to the work performed?
  • Is the Article 16(3) activity and identity statement on the device, or, where that is impracticable, on the packaging or an accompanying document?
  • Do the source and local revisions match the approved translation relationship?
  • Is affected stock identifiable through the job and subsequent stock movements?
  • Are execution results resolved, with supporting decisions traceable?
  • Does the market file match the intended destination and device scope?
  • Is the human disposition recorded, including the decision-maker and stock covered?

Treat broken associations as findings with an owner and closure reviewer, record whether each blocks distribution, and carry that status into device market-release review.

Frequently asked questions

Q.How should a translation-supplier change be assessed when the approved wording remains unchanged?+

Treat the supplier replacement as a process change rather than assuming unchanged wording means unchanged risk. A practical change-control assessment compares translator competence, terminology controls, source-file access, subcontracting, proofreading and escalation arrangements. Confirm that the approved text, layout and revision identifiers remain unchanged in the delivered files. Record whether supplier qualification or additional review is appropriate, and who accepts the outcome. Keep that decision separate from any assessment of notification implications.

Q.Does every repeat batch need a fresh notification, and how does non-legally-binding MDCG guidance address that question?+

No, according to non-legally-binding MDCG 2021-26, Question 7. It notes there are no specific requirements in the Regulations, and considers that notification is only due when a device is going to be relabelled and/or repackaged for the first time; the notification does not have to be repeated on a unit-by-unit or batch-by-batch basis. Operationally, retain a comparison showing that repeat work remains within the previously notified scope; Question 8's triggers (next FAQ) define when that scope has changed.

Q.When should a new device, language, packaging change or certificate change trigger a notification review?+

For EU medical devices, non-legally-binding MDCG 2021-26, Question 8 recommends competent-authority notification for previously unnotified devices or languages, modified accompanying information or outer packaging, certificate changes including renewal, and a change of notified body. Use these events as notification-review triggers before implementing the change. Assess manufacturer communications separately rather than assuming identical recipient criteria. A practical device market-release review records the applicable trigger, affected stock and the person responsible for resolving notification actions.

Q.How should a sample or mock-up request be handled when the planned availability date is approaching?+

For EU medical devices, non-legally-binding MDCG 2021-26, Question 6 recommends providing the competent authority, on request, a sample or mock-up including translated labelling and instructions for use. Operationally, log the request, assign an owner and identify the exact configuration requested. Preserve the submitted version and dispatch evidence. If it differs from planned stock, escalate the discrepancy through the QA disposition step; reconsider availability rather than substituting an obsolete specimen to meet the schedule.

Q.What should be considered when relabelling stops but labelled stock and customer records remain?+

For EU medical devices, non-legally-binding MDCG 2021-26, Question 9 suggests informing the manufacturer and concerned competent authority of cessation because this information could support field safety corrective action processes. As a practical closure measure, distinguish stopping operations from disposing of remaining stock. Assign ownership for customer enquiries, complaints, revision histories and distribution records. Maintain usable contact routes and recall readiness, and document how residual stock will be assessed before further supply.

Q.How should mixed-language kits be reviewed when their components reference different source revisions?+

Different revision identifiers are not, by themselves, proof of incompatibility. A practical medical device kitting review maps each component to its manufacturer source, translated version, language and intended use within the kit. Examine cross-references, warnings, assembly instructions and shared accessories for contradictions. Seek manufacturer clarification where the relationship between revisions is uncertain. Record the accepted configuration and rationale; avoid replacing every component’s instructions with the newest-looking document without checking applicability.

Q.How should a returned device with an obsolete translated IFU be assessed before redistribution?+

Keep the return separate from distributable stock while assessing whether the translated instructions remain applicable to that particular device. A practical returns quarantine assessment examines device identity, manufacturer source revision, reasons for supersession, packaging condition and return history. Determine whether replacement instructions alone address the issue or whether other investigation is needed. Record any controlled replacement and reconciliation of removed material. A designated human reviewer should decide disposition rather than allowing the return transaction to restore saleable status.

Q.What should the reviewer do when notification receipt evidence or notified-body certificate scope is unclear?+

Separate uncertainty about communication from uncertainty about certification. For notification records, examine the recipient, submission contents, dispatch evidence and any acknowledgement; seek clarification rather than inventing receipt dates. For certificate scope, compare the actual activity and device coverage with the certificate and obtain written clarification where needed. Keep affected stock on hold while either question is open.

Primary sources

  • Regulation (EU) 2017/745 (MDR), consolidated text 9 November 2026, Article 16
  • MDCG 2021-26 Q&A on repackaging & relabelling activities under Article 16 MDR/IVDR (October 2021)
  • MDCG 2021-23 Guidance for notified bodies, distributors and importers on certification activities per Article 16(4) (August 2021)
Software that covers Device Relabelling
V5 Ultimate Medical Device MES
Device history records built from signed steps at the workstation: component lots, configured step order, inspections, UDI and…
V5 Ultimate for EU MDR
V5 assembles it as one file that a Notified Body auditor can navigate in an hour.
Talk to us about Device Relabelling

Want to see how Device Relabelling could fit into your own records and workflows? Explore the related V5 pages or talk to our team about what applies to your operation.

Start free
Back to glossary
Where this term comes up
Medical Devices
Related terms
  • → UDI
  • → Own-Label Distributor
  • → Artwork Proof Approval
  • → EU MDR
  • → Initial Importer

Next Step

Try V5 with your own records, or ask a question first. Ask V5 opens with an editable question; nothing is sent until you choose to.

Start your free trial Browse All Features

Showing the site for the United States. Local spelling, dates and prices are applied automatically; change the language at the top of the page.

V5 Ultimate
Ultimate

Warehouse, quality and manufacturing software for regulated operations.

ProductIndustriesPricingResourcesSecurity & TrustCompanyLegal Centre

© S.G. Systems, LLC