V5 Ultimate
Ultimate
PricingResourcesCompany
English (original)العربيةDeutschΕλληνικάEspañolFrançaisItalianoעברית日本語한국어NederlandsPolskiPortuguêsSvenska
Talk to us Log in Start free trial
Food & Beverage
  • Bakery & Confectionery
  • Food Processing
  • Ingredients & Dry Mixes
  • Sausage & Meat Products
  • Produce Packing
  • Pet Food & Pet Supplements
  • Dairy & Chilled Foods
  • Beverages & Liquid Manufacturing
  • Infant Formula
Pharma & Life Sciences
  • Radiopharmaceuticals
  • Pharmaceutical
  • Medical Devices
  • Veterinary Pharmaceuticals
  • Blood & Tissue
  • Diagnostics & IVD Kits
  • Reference Materials & Standards
  • OTC Topical Creams & Gels
Supplements & Personal Care
  • Dietary Supplements
  • Cosmetics & Personal Care
  • Cannabis & Hemp Manufacturing
  • Fragrances & Perfumery
  • Soaps & Personal Hygiene
  • Botanical Extracts & Essential Oils
Chemicals & Materials
  • Plastics & Resins
  • Agricultural Chemicals
  • Specialty & Industrial Chemicals
  • Printing Inks, Pigments & Construction Chemicals
  • Lubricants & Metalworking Fluids
  • Coatings, Adhesives & Sealants
Other Manufacturing
  • Consumer Products
  • Electronics & Electrical Assembly
  • Textiles & Nonwoven
  • Contract Manufacturing & Co-Packing
  • Packaging & Food-Contact Materials
  • Regulated Warehousing & Distribution
Explore all industries
  • Warehouse & TraceabilityReceive, store, pick and trace
  • Production & PlanningPlan jobs and guide production
  • Quality & LaboratoryTest, investigate and review
  • Documents & TrainingControl procedures and qualifications
  • Equipment & PerformanceMaintain assets and measure losses
  • Orders & IntegrationsConnect customers, suppliers and systems
  • Outputs & InsightsReports, KPIs and label design
Why V5? Product & Features
Why V5?Product & Features
  • Warehouse & TraceabilityReceive, store, pick and trace
  • Production & PlanningPlan jobs and guide production
  • Quality & LaboratoryTest, investigate and review
  • Documents & TrainingControl procedures and qualifications
  • Equipment & PerformanceMaintain assets and measure losses
  • Orders & IntegrationsConnect customers, suppliers and systems
  • Outputs & InsightsReports, KPIs and label design
Explore all industries
    • Bakery & Confectionery
    • Food Processing
    • Ingredients & Dry Mixes
    • Sausage & Meat Products
    • Produce Packing
    • Pet Food & Pet Supplements
    • Dairy & Chilled Foods
    • Beverages & Liquid Manufacturing
    • Infant Formula
    • Radiopharmaceuticals
    • Pharmaceutical
    • Medical Devices
    • Veterinary Pharmaceuticals
    • Blood & Tissue
    • Diagnostics & IVD Kits
    • Reference Materials & Standards
    • OTC Topical Creams & Gels
    • Dietary Supplements
    • Cosmetics & Personal Care
    • Cannabis & Hemp Manufacturing
    • Fragrances & Perfumery
    • Soaps & Personal Hygiene
    • Botanical Extracts & Essential Oils
    • Plastics & Resins
    • Agricultural Chemicals
    • Specialty & Industrial Chemicals
    • Printing Inks, Pigments & Construction Chemicals
    • Lubricants & Metalworking Fluids
    • Coatings, Adhesives & Sealants
    • Consumer Products
    • Electronics & Electrical Assembly
    • Textiles & Nonwoven
    • Contract Manufacturing & Co-Packing
    • Packaging & Food-Contact Materials
    • Regulated Warehousing & Distribution
PricingResourcesCompany
Start free trial Log in Talk to us
HomeGlossaryInvestigational Plan
Compliance · The complete guide

Investigational Plan

In short

Investigational Plan is the sponsor document required for a US device IDE that sets out, under 21 CFR 812.25, the investigation's purpose, protocol, risk analysis, device description, monitoring, labeling, consent materials, IRB and institution information and additional records.

4,520 words · ~21 min read
A clinical affairs specialist reviews a tabbed investigational plan binder beside a tablet change log and a boxed investigational device prototype.
The investigational plan version governs which device configuration is supplied; plan changes are reviewed through their own route.
On this page
  1. 01Investigational Plan
  2. 02Separate the plan, protocol, and IDE application
  3. 03Review the plan's contents and linked evidence
  4. 04Identify the plan baseline that covers the device
  5. 05Classify a change before updating the baseline
  6. 06Carry the approved baseline through supply
  7. 07Connect versions, authorizations, and shipment records
  8. 08Work through plan, device, and evidence changes
  9. 09Recover from broken links and conflicting records
  10. 10Check readiness for a plan-based supply decision
On this page · 10 sections
  1. 1Investigational Plan
  2. 2Separate the plan, protocol, and IDE application
  3. 3Review the plan's contents and linked evidence
  4. 4Identify the plan baseline that covers the device
  5. 5Classify a change before updating the baseline
  6. 6Carry the approved baseline through supply
  7. 7Connect versions, authorizations, and shipment records
  8. 8Work through plan, device, and evidence changes
  9. 9Recover from broken links and conflicting records
  10. 10Check readiness for a plan-based supply decision
AI · Explain it for MY operation

How does Investigational Plan apply to your shop floor?

Pick your industry and scale — Ask V5 rewrites the definition in your context, gives a worked example, and shows what V5 does on day one.

Your scale

01Investigational Plan

What an investigational plan brings together

Investigational Plan is the documented framework for a US investigational device study, bringing its protocol, device information, and oversight materials together. The aliases IDE investigational plan and 21 CFR 812.25 investigational plan refer to this same subject.

This article addresses US FDA device investigational device exemption (IDE) documentation and manufacturer configuration control. Its central manufacturer question is: which documented plan baseline covers the device configuration proposed for supply? The practical focus is the connection between study documentation and the physical device—not simply whether a document bearing the plan title exists.

For US device IDE applicability, consult the current definitions in 21 CFR 812.3. Use the significant risk device overview for adjacent applicability context, not as a classification decision for a particular device. This article does not determine device risk classification or assume that every device investigation follows the same regulatory pathway.

The sections that follow distinguish the documents, map plan contents, and examine version and change decisions from the manufacturer's perspective. They support a structured review by the people responsible for applicability and supply disposition; they do not make those decisions. Other markets, drug investigations, clinical decisions, treatment, and study-design advice are outside scope. This article is not legal advice.

02Separate the plan, protocol, and IDE application

For US device IDE applications, FDA’s IDE Application checklist presents the protocol within its investigational-plan checklist. The protocol is therefore a component of the plan, not a synonym for the complete plan. When reviewing a document, distinguish its scope from its title.

Document or termScope in this articleManufacturer decision
Investigational planThe broader plan containing the protocol and other plan material.Identify the complete document baseline being reviewed.
Clinical protocolA component addressing how the investigation is conducted.Locate its version within the plan baseline.
Protocol amendmentA document-change term for a revision to the protocol component.Identify affected documents; do not infer an FDA change classification from the title.
IDE applicationThe submission containing plan material and application-level information.Distinguish the submitted package from its individual documents.
ISO 14155 clinical investigation planA neighboring standards topic: ISO 14155 clinical investigation.Keep the standards document context separate; no equivalence assessment is made here.

For US device IDE applications, FDA’s IDE Application page also lists a description of the methods, facilities, and controls used for manufacturing, processing, packing, storage, and installation of the device. This application-level material illustrates why neither the protocol nor the plan should be treated as interchangeable with the application.

In controlled-document practice, a protocol amendment commonly identifies a changed component rather than a replacement for the entire document baseline. A useful review question is: which component changed, and where is that revision connected to the broader plan? Record that relationship explicitly instead of relying on matching filenames or revision dates.

03Review the plan's contents and linked evidence

For US device IDE investigations, 21 CFR 812.25 states that the investigational plan shall include ten items, in a stated order, as paragraphs (a) through (j). The table paraphrases each paragraph of the current regulation closely; the right-hand column is a proposed manufacturer review question, not an additional legal requirement.

The evidence references and manufacturer questions are proposed review practice, not additional legal content requirements. Use them to identify the document owner, referenced revision, affected device configuration, and any unresolved mismatch. Review the protocol and its supporting analysis as document contents here, not as instructions for designing a study.

ParagraphRegulationWhat the current text requiresManufacturer review question
(a) Purpose21 CFR 812.25(a)The name and intended use of the device and the objectives and duration of the investigation.Who owns this statement, and does its device identity match the configuration being reviewed?
(b) Protocol21 CFR 812.25(b)A written protocol describing the methodology to be used and an analysis of the protocol demonstrating that the investigation is scientifically sound.Which protocol revision is referenced, who owns it, and is its supporting analysis linked without ambiguity?
(c) Risk analysis21 CFR 812.25(c)A description and analysis of all increased risks to which subjects will be exposed, the manner in which these risks will be minimized, a justification for the investigation, and a description of the patient population, including the number, age, sex, and condition.Which analysis revision addresses the described device, and who resolves a mismatch between that analysis and the configuration?
(d) Description of device21 CFR 812.25(d)A description of each important component, ingredient, property, and principle of operation of the device and of each anticipated change in the device during the course of the investigation.Can the technical owner identify the covered configuration and distinguish it from descriptions of anticipated changes?
(e) Monitoring procedures21 CFR 812.25(e)The sponsor's written procedures for monitoring the investigation and the name and address of any monitor.Who owns these references, and are conflicting procedure revisions or monitor details unresolved?
(f) Labeling21 CFR 812.25(f)Copies of all labeling for the device.Does the labeling owner identify a complete set for the configuration, including any conflicting revision references?
(g) Consent materials21 CFR 812.25(g)Copies of all forms and informational materials to be provided to subjects to obtain informed consent.Who owns the set, and do device descriptions in these materials conflict with the referenced configuration?
(h) IRB information21 CFR 812.25(h)A list of the names, locations, and chairpersons of all IRBs that have been or will be asked to review the investigation, and a certification of any action taken by any of those IRBs with respect to the investigation.Who maintains the listing, and can the reviewer identify which document revisions the associated action concerns?
(i) Other institutions21 CFR 812.25(i)The name and address of each institution at which a part of the investigation may be conducted that is not identified under paragraph (h).Who owns this listing, and are inconsistent institution names or addresses clearly identified for resolution?
(j) Additional records and reports21 CFR 812.25(j)A description of records and reports that will be maintained on the investigation in addition to those prescribed in subpart G.Who owns each named record or report, and can the reviewer locate the referenced version rather than an unrelated template?

Two items are easy to under-read. Paragraph (d) covers anticipated device changes as well as the current design, so the plan can name expected revisions before they occur. Paragraph (h) asks for a certification of any action an IRB has taken, not just a list of IRB names; a review should therefore be able to locate the certification that corresponds to each listed IRB.

04Identify the plan baseline that covers the device

For a manufacturer supplying devices for a US FDA IDE investigation, an approved baseline is a useful operational description of identified document versions and their recorded authorizations—not a new FDA-defined term. The practical question is which baseline covers the configuration proposed for supply to the intended destination, rather than which plan file was most recently uploaded.

A practical comparison chain connects the plan revision, its referenced protocol revision, the device configuration, the labeling revision, the destination scope, and any authorization conditions. A linked configuration specification can identify the physical configuration being compared. Record each relationship explicitly; matching the plan title alone leaves the remaining scope unresolved.

Example document-control states distinguish draft, approved, effective, and superseded. Draft identifies work still under review; approved records a completed approval within its stated scope; effective identifies the version designated for use under applicable conditions; superseded identifies a version replaced within a defined scope. These are illustrative local states, not a universal FDA workflow. A newer file does not automatically replace every earlier baseline for every destination.

For review clarity, keep the following record purposes separately identifiable rather than combining them into one approval flag:

  • FDA action: the agency action, its documented scope, and any stated conditions.
  • IRB action: the board’s documented review outcome and the investigation materials and site scope it addresses.
  • Sponsor authorization: the sponsor’s documented direction about which baseline is authorized for the proposed activity.
  • Manufacturer supply disposition: the accountable decision about the identified devices proposed for supply.

This separation does not prescribe a universal approval sequence. Accountable people determine applicability and disposition using the relevant records and conditions. An identifier match supports that review; it is not the decision itself. An electronic release record can serve as a record location for the resulting disposition and its baseline references. Amendment classification belongs to the following section; operational handoffs belong to the lifecycle section.

05Classify a change before updating the baseline

A practical change review starts with the actual difference, not the amendment title. A protocol amendment, device revision, or administrative update identifies what changed; it does not establish the regulatory route.

Start from the prior-approval rule

Under 21 CFR 812.35(a)(1), except as described in paragraphs (a)(2) through (a)(4), a sponsor must obtain approval of a supplemental application under § 812.30(a), and IRB approval when appropriate, prior to implementing a change to an investigational plan. Prior approval is therefore the default; the other routes are exceptions that must each be shown to apply.

Emergency deviations are a separate category

Paragraph (a)(2) removes the supplement-approval requirement for a deviation from the investigational plan to protect the life or physical well-being of a subject in an emergency. Such a deviation is reported to FDA within 5 working days after the sponsor learns of it. It is not a planned change route and should not be used to classify a revision.

Five-day notice: developmental and protocol changes

Paragraph (a)(3) allows certain changes without prior approval if the sponsor determines, on the basis of credible information defined in (a)(3)(iii), that they meet the criteria in (a)(3)(i) or (a)(3)(ii), and provides notice to FDA within 5 working days of making them.

  • 812.35(a)(3)(i): developmental changes in the device, including manufacturing changes, that do not constitute a significant change in design or basic principles of operation and that are made in response to information gathered during the investigation.
  • 812.35(a)(3)(ii)(A): clinical-protocol changes that do not affect the validity of the data or information resulting from completion of the approved protocol, or the relationship of likely patient risk to benefit relied upon to approve it.
  • 812.35(a)(3)(ii)(B): changes that do not affect the scientific soundness of the investigational plan.
  • 812.35(a)(3)(ii)(C): changes that do not affect the rights, safety, or welfare of the human subjects.

The notice clock under 812.35(a)(3)(iv) starts from a defined event, not from approval of an internal document. Device changes are deemed to occur on the date the device manufactured with the design or manufacturing change is distributed to the investigators. Protocol changes are deemed to occur when the sponsor notifies a clinical investigator that the change should be implemented, or, for sponsor-investigator studies, when the change is incorporated in the protocol. For a manufacturer, the shipment record of the first changed unit is therefore part of the change evidence.

Credible information has route-specific content under 812.35(a)(3)(iii). For developmental device changes, it includes data generated under the design and development activities of §820.10(c), preclinical or animal testing, peer-reviewed published literature, or other reliable information such as clinical information gathered during a trial or marketing. For protocol changes, it is the sponsor's documentation supporting the conclusion that the change has no significant impact on study design or planned statistical analysis and does not affect subjects' rights, safety or welfare. That documentation includes information such as peer-reviewed literature, investigator recommendations or trial or marketing data. A document marked approved is not itself this supporting information.

Annual report: minor changes with the same no-effect test

Paragraph (a)(4) applies to minor changes to the purpose of the study, risk analysis, monitoring procedures, labeling, informed consent materials, and IRB information that do not affect the three matters listed in (a)(4)(i) through (iii) — the same data-validity and risk-benefit, scientific-soundness, and subject-welfare criteria described above. Such changes are reported in the annual progress report for the IDE under § 812.150(b)(5). Membership in the list alone does not establish eligibility; the no-effect determination does.

Record a decision that can be followed

A useful change-control record states the actual difference, the route considered, the credible information relied upon, the deemed-occurrence event and date where a notice applies, and the plan, protocol and device revisions before and after. Whether a specific change qualifies is a sponsor regulatory decision; the manufacturing record documents it rather than making it.

06Carry the approved baseline through supply

For a US device IDE investigation, the following is an illustrative operating model, not a statutory allocation of responsibilities. Sponsor regulatory assessment, engineering configuration identification, clinical-operations document coordination, and manufacturer supply disposition remain separate decisions. At each handoff, an unanswered acceptance question keeps that handoff open.

Document transfer does not establish device movement. Supply authorization, dispatch, and receipt are distinct events; reconciliation connects their records rather than treating any one as proof of the others.

From plan documents to supply disposition
  1. 1
    AssembleIdentify plan components and their controlled versions.
  2. 2
    CompareMatch the proposed device configuration to the baseline.
  3. 3
    AssessDocument differences, evidence, and the route decision.
  4. 4
    AuthorizeRecord applicable decisions, conditions, and responsible people.
  5. 5
    ActivateIdentify effective versions and their destination scope.
  6. 6
    Review supplyRecord the human disposition for the proposed shipment.
  7. 7
    ReconcileConnect shipment records, receipt information, and later changes.
Illustrative manufacturer controls connect document decisions with device supply.

07Connect versions, authorizations, and shipment records

Use this proposed record design to connect the investigational plan to manufacturer records. It is a traceability practice, not a mandated system, universal record format, or software capability.

Plan [document control]: plan ID, revision, evidence references ↕ explicit revision relationship Configuration [engineering]: device ID, configuration revision ↕ applicability relationship Authorization [clinical operations]: decision ID, plan revision, configuration revision, investigator/site ID, decision maker, date, evidence versions considered ↓ decision reference Shipment [supply]: shipment ID, device identifiers, configuration revision, destination, authorization ID

Treat the named records as authoritative for their respective fields. Assign a named document owner, engineering configuration owner, clinical-operations decision owner, and supply-record owner to resolve conflicting values in their domains. A copied plan revision in a shipment record remains a reference; it does not transfer document ownership. Record the source identifier alongside each copied value. These relationships support data integrity by design.

  • Which stable identifier crosses each boundary, and which receiving field stores it?
  • Which exact revision does the reference identify?
  • When was the receiving record reconciled, by whom, and against which source?
  • How is an unresolved match recorded, including its status and responsible owner?

Apply these questions to manual transfers as well as manufacturing–quality system integrations. Preserve each decision’s links to the evidence versions considered at that time. Later revisions can have separate relationships, but should not silently replace historical references. This keeps the decision reconstructable without treating the latest document as evidence of an earlier authorization.

Records connecting plan and configuration

Plan baseline

Identifies the document set under review.

  • plan_id
  • plan_revision
  • protocol_revision
  • effective_scope

Device configuration

Identifies the device and associated labeling.

  • device_id
  • device_revision
  • configuration_reference
  • labeling_revision

Change record

Connects a proposed difference to its assessment.

  • change_id
  • affected_revisions
  • evidence_reference
  • route_status

Authorization

Preserves the issuer, scope, and conditions.

  • authorization_id
  • issuer
  • authorized_scope
  • conditions

Supply disposition

Links a shipment to its reviewed baseline.

  • shipment_id
  • plan_revision
  • device_revision
  • site_id
  • approved_by
A proposed evidence model, independent of any particular software.

08Work through plan, device, and evidence changes

These fictional US device IDE cases illustrate manufacturer record decisions, not regulatory classifications. IP-R2 and D3 are version identifiers, not quantities or risk categories. Each disposition is bounded by the stated assumptions; none provides clinical, study-design or legal advice.

Fictional case: IP-R2 versus a protocol amendment

  • Assumptions: The sponsor proposes plan revision IP-R2, whose revised index references an amended protocol. The plan and protocol retain separate document identities and revision histories.
  • Conflicting record: The change request names only the protocol amendment, while the proposed plan index points to that amendment without linking their review records.
  • Review question: Send the linked documents for change-control review: what changed, which route applies, and what plan-version action follows under the sponsor’s document controls?
  • Bounded disposition: Keep IP-R2 proposed until the responsible reviewers resolve its status. Do not infer that every protocol amendment requires the same plan-version action.
  • Traceability retained: Record the plan-to-protocol relationship, changed index entry, component revision, change request and resulting decision separately, preserving the prior relationship.

Fictional case: D3 supplied under IP-R2

  • Assumptions: Approved plan IP-R2 explicitly covers device configuration D3 for the destination. This fictional manufacturer’s procedure requires confirmation before shipment.
  • Conflicting record: The proposed shipment references an older plan baseline, although the device identity and configuration specification identify D3.
  • Review question: Does the shipment’s actual configuration and destination match IP-R2’s approved scope, and is the older reference merely a record error?
  • Bounded disposition: Reconcile the shipment reference to IP-R2 after checking the underlying records. An authorized human records the shipment disposition before D3 is shipped; this sequence follows the assumed procedure, not a universal FDA shipment rule.
  • Traceability retained: Preserve the original reference, correction reason, IP-R2 scope reference, D3 identity, destination and signed disposition with the shipment record.

Fictional case: Change-record evidence pending

  • Assumptions: A proposed configuration change has an incomplete supporting assessment and no completed change-route decision.
  • Incomplete record: The draft configuration exists, but its assessment lacks the information needed for the sponsor’s route review.
  • Review question: What evidence remains outstanding, who will provide it, and who owns the implementation decision?
  • Bounded disposition: Mark evidence pending, assign an assessment owner and leave implementation unresolved. Do not promote the changed configuration into the supply baseline or automatically classify the change as requiring a supplement.
  • Traceability retained: Link the proposal, missing assessment items, assigned owner and pending decision to the unchanged supply baseline. Record the eventual decision separately when review concludes.

Fictional manufacturing case: IP-M42 configuration transition

In fictional IP-M42, P1 covers C2 and label L2 at Site North. A connector revision C3 follows assembly observations during the investigation. No significance determination is supplied. A17 remains open; P2 is draft. Under the assumed local procedure, changed-unit dispatch is unresolved.

Controlled rework can leave different configurations within one lot. Manufacturing checks each serial’s build and rework records before assigning configuration. These illustrative traceability controls are not a prescribed FDA form.

Physical unitsVerified configurationPlan relationshipProposed disposition
Lot M42 / serials 101–106C2; label L2P1 covers C2 at Site NorthSeparate unchanged-unit review against P1
Lot M42 / serials 107–110C3; label L3 proposedP2 draft; A17 openChanged-unit dispatch remains unresolved
Lot M43 / serials 111–114C3 awaiting completed build reconciliationNo completed P2 supply relationshipResolve identity as well as assessment before disposition

Drawing approval does not approve C3 for investigational supply. A17 records the difference, investigation information prompting it and missing evidence. The sponsor reviews §812.35(a)(1) and (a)(3): prior approval is the default; developmental-change notice is conditional. Calling the connector change minor does not establish its eligibility.

Branch one assumes the sponsor documents eligibility under §812.35(a)(3)(i), with credible information under (a)(3)(iii)(A); connectors are not generally classified here. Link the determination to P2’s description and labeling. Under the assumed local procedure, Site North acknowledges the package before dispatch. Receipt is not FDA approval or any applicable IRB action.

Disposition permits shipment of serials 107–110 through dispatch D42, recording C3 units, labeling, destination, plan and decision references. The sponsor receives the actual distribution date. Under §812.35(a)(3)(iv), device changes occur when changed devices are distributed to investigators. Retain dispatch evidence for the notice clock rather than the drawing-approval, upload or acknowledgment date, with the notice and supporting information separately identifiable.

Branch two assumes A17 cannot establish exception criteria. The sponsor considers prior approval and records remaining actions. Changed serials stay outside the assumed dispatch authorization until applicable decisions are established. Serials 101–106 receive separate review against P1, destination scope and conditions; a shared lot number neither approves nor stops every unit.

Retain the crosswalk, build/rework references, A17 evidence, sponsor determination, applicable agency/IRB records, site acknowledgment and signed disposition. Reconcile shipped serials without silently revising the original list. A C3 decision must not rewrite the baseline supporting earlier C2 shipments. Manufacturing evidence supports, not replaces, investigation decisions.

09Recover from broken links and conflicting records

These practice-based controls focus recovery on affected records and device configurations, rather than assigning every discrepancy the same supply consequence.

  • Authorization attached to the wrong revision: compare the authorization’s referenced revision with the linked plan. Sponsor regulatory affairs resolves whether it supports the proposed supply decision. Close with the correct authorization, corrected linkage, and documented assessment of affected configurations.
  • Destination scope copied without review: compare the destination against the applicable plan baseline. The sponsor’s investigation lead resolves destination eligibility before further dispatch there. Close with reviewed scope evidence and the recorded destination decision, not merely a corrected address.
  • Historical baseline overwritten: check shipment-time references against version history. Manufacturer document control restores the historical record; quality assesses the shipment-to-baseline conclusion. Close with the preserved original, recoverable version history, and a documented reconciliation that protects data integrity.
  • Cross-system identifier match failed: compare plan, device, and shipment identifiers across the affected systems. The integration owner reconciles identifiers; manufacturer quality resolves configuration identity for supply review. Close with source records, corrected mapping, and a successful reconciliation check across the connected systems.

If discovered after shipment, preserve original records and escalate shipment, configuration, destination, and authorization facts to responsible sponsor and manufacturer functions. Their assessment determines disposition; the discrepancy alone does not establish recall, continued use, or a regulatory reporting outcome.

10Check readiness for a plan-based supply decision

Use these proposed review questions before deciding whether to supply a device under the identified plan baseline:

  • Document identity: Which controlled record identifies the applicable plan and revision?
  • Component versions: Where does the plan index identify each referenced component’s revision?
  • Configuration coverage: Which configuration assessment connects the device being supplied to that baseline?
  • Destination scope: Which record establishes coverage for the intended investigator and site?
  • Authorization conditions: Where are applicable conditions recorded, and who checks their status?
  • Change route: Which sponsor decision records the route and outstanding actions?
  • Evidence completeness: Which evidence register identifies missing references and their owners?
  • Accountable disposition: Who records the supply decision and resolves open exceptions?
  • Historical traceability: Can the shipment reference recover the baseline and decision applicable at dispatch?

Record each answer separately from its supporting reference. Keep unresolved answers assigned to an owner with a disposition status for exception review. Use audit trail review to examine disputed record changes. Checklist completion does not certify compliance.

Frequently asked questions

Q.Who reconciles the plan baseline when the IDE sponsor and device manufacturer are different organizations?+

A practical arrangement is sponsor-led reconciliation, with the manufacturer confirming the device configuration and associated manufacturing documentation. The organizations can assign named owners for exchanging plan revisions, resolving discrepancies, and acknowledging the baseline applicable to supply. Their agreement should distinguish sponsor decisions about the investigation from manufacturer decisions about product disposition. A shared change-control process helps connect those decisions without assuming that either organization’s document approval automatically updates the other’s controlled records.

Q.Does an IRB document alone establish which device configuration is covered for supply?+

For supply reconciliation, an IRB document is best treated as part of the investigation’s documentation, rather than a standalone configuration specification. The practical check is whether the applicable plan baseline, device description, change decision, and manufacturer’s configuration record identify the same device. If those records disagree, a prudent approach is to hold the affected supply decision for human resolution. A controlled configuration specification provides the detailed product identity against which that reconciliation can be performed.

Q.How can a site identify the labeling and consent-material versions associated with its plan baseline?+

A useful document-control practice is to give the site a baseline index identifying each applicable labeling item and consent material by document identifier, revision, and effective status. The index can distinguish sponsor-controlled materials from site-specific versions and identify who resolves mismatches. Staff can then compare materials available for use against that index rather than relying on filenames or email dates. Keep the investigational-device-labeling reference connected to the actual controlled documents, not merely a generic labeling template.

Q.Can one plan baseline be linked to more than one device configuration?+

As a configuration-management approach, a baseline index can map a plan to multiple device configurations; the mapping itself does not establish that each configuration is covered for supply. Give each configuration a distinct identity and document the basis for associating it with that baseline, including any site or implementation restrictions. Avoid an unrestricted list of interchangeable revisions. Connecting this mapping to investigational device accountability helps reviewers identify which configuration was actually sent to each investigator.

Q.Does describing an anticipated device change resolve its change-control classification?+

For practical review, separate the description of an anticipated change from the documented decision about implementing the actual change. The description provides context; it should not serve as a substitute for evaluating the specific proposal and its supporting information. If change-record evidence is pending, record that unresolved status rather than treating anticipation as authorization. Linking the review to relevant design-control records helps distinguish what was originally contemplated from what was developed, evaluated, and presented for a human decision.

Q.What does a protocol-only amendment mean for the plan's document index?+

Refer to the protocol-amendment case above for the approval decision; the indexing task is to identify which components actually changed. A practical amendment index distinguishes replaced attachments, unchanged references and newly introduced material. Check page-level identifiers against the index before distributing the package, and identify its issuer and intended recipients. Keep the associated audit trail linked to the index so an attachment-order correction remains distinguishable from a substantive protocol revision.

Q.How should a superseded plan version remain traceable after another baseline becomes effective?+

The historical-link rule is explained in the evidence-model section. The archive task is to test whether someone outside the original project can retrieve the cited package. A practical retrieval check starts from a shipment identifier, follows its production context tagging references, and opens the corresponding plan index and attachments. Record missing files, inaccessible locations and unreadable exports separately; assign their recovery to the record owner rather than substituting a current package for the missing historical one.

Q.How should a cross-market investigation use this US FDA-focused explanation?+

Use this explanation as a US FDA-focused reference for document and supply reconciliation, not as a cross-market authorization checklist. Other markets are outside its scope. A practical cross-market matrix can identify the shared investigation documents, jurisdiction-specific materials, and the people responsible for determining local applicability. Consult the separate ISO 14155 clinical investigation overview for that standard’s general purpose rather than treating its terminology as interchangeable with the US investigational plan. This explanation is not legal, clinical, treatment, or study-design advice.

Primary sources

  • 21 CFR 812.25 Investigational plan
  • 21 CFR 812.35 Supplemental applications
  • FDA IDE Application
Talk to us about Investigational Plan

Want to see how Investigational Plan could fit into your own records and workflows? Explore the related V5 pages or talk to our team about what applies to your operation.

Start free
Back to glossary
Where this term comes up
Medical Devices
Related terms
  • → Investigational Device Labeling
  • → Investigational Device Accountability
  • → Significant Risk Device
  • → ISO 14155

Next Step

Try V5 with your own records, or ask a question first. Ask V5 opens with an editable question; nothing is sent until you choose to.

Start your free trial Browse All Features

Showing the site for the United States. Local spelling, dates and prices are applied automatically; change the language at the top of the page.

V5 Ultimate
Ultimate

Warehouse, quality and manufacturing software for regulated operations.

ProductIndustriesPricingResourcesSecurity & TrustCompanyLegal Centre

© S.G. Systems, LLC