V5 Ultimate
Ultimate
PricingResourcesCompany
English (original)DeutschΕλληνικάEspañolFrançaisItalianoעברית日本語한국어NederlandsPolskiPortuguêsSvenska
Talk to us Log in Start free trial
Food & Beverage
  • Bakery & Confectionery
  • Food Processing
  • Ingredients & Dry Mixes
  • Sausage & Meat Products
  • Produce Packing
  • Pet Food & Pet Supplements
  • Dairy & Chilled Foods
  • Beverages & Liquid Manufacturing
  • Infant Formula
Pharma & Life Sciences
  • Radiopharmaceuticals
  • Pharmaceutical
  • Medical Devices
  • Veterinary Pharmaceuticals
  • Blood & Tissue
  • Diagnostics & IVD Kits
  • Reference Materials & Standards
  • OTC Topical Creams & Gels
Supplements & Personal Care
  • Dietary Supplements
  • Cosmetics & Personal Care
  • Cannabis & Hemp Manufacturing
  • Fragrances & Perfumery
  • Soaps & Personal Hygiene
  • Botanical Extracts & Essential Oils
Chemicals & Materials
  • Plastics & Resins
  • Agricultural Chemicals
  • Specialty & Industrial Chemicals
  • Printing Inks, Pigments & Construction Chemicals
  • Lubricants & Metalworking Fluids
  • Coatings, Adhesives & Sealants
Other Manufacturing
  • Consumer Products
  • Electronics & Electrical Assembly
  • Textiles & Nonwoven
  • Contract Manufacturing & Co-Packing
  • Packaging & Food-Contact Materials
  • Regulated Warehousing & Distribution
Explore all industries
  • Warehouse & TraceabilityReceive, store, pick and trace
  • Production & PlanningPlan jobs and guide production
  • Quality & LaboratoryTest, investigate and review
  • Documents & TrainingControl procedures and qualifications
  • Equipment & PerformanceMaintain assets and measure losses
  • Orders & IntegrationsConnect customers, suppliers and systems
  • Outputs & InsightsReports, KPIs and label design
Why V5? Product & Features
Why V5?Product & Features
  • Warehouse & TraceabilityReceive, store, pick and trace
  • Production & PlanningPlan jobs and guide production
  • Quality & LaboratoryTest, investigate and review
  • Documents & TrainingControl procedures and qualifications
  • Equipment & PerformanceMaintain assets and measure losses
  • Orders & IntegrationsConnect customers, suppliers and systems
  • Outputs & InsightsReports, KPIs and label design
Explore all industries
    • Bakery & Confectionery
    • Food Processing
    • Ingredients & Dry Mixes
    • Sausage & Meat Products
    • Produce Packing
    • Pet Food & Pet Supplements
    • Dairy & Chilled Foods
    • Beverages & Liquid Manufacturing
    • Infant Formula
    • Radiopharmaceuticals
    • Pharmaceutical
    • Medical Devices
    • Veterinary Pharmaceuticals
    • Blood & Tissue
    • Diagnostics & IVD Kits
    • Reference Materials & Standards
    • OTC Topical Creams & Gels
    • Dietary Supplements
    • Cosmetics & Personal Care
    • Cannabis & Hemp Manufacturing
    • Fragrances & Perfumery
    • Soaps & Personal Hygiene
    • Botanical Extracts & Essential Oils
    • Plastics & Resins
    • Agricultural Chemicals
    • Specialty & Industrial Chemicals
    • Printing Inks, Pigments & Construction Chemicals
    • Lubricants & Metalworking Fluids
    • Coatings, Adhesives & Sealants
    • Consumer Products
    • Electronics & Electrical Assembly
    • Textiles & Nonwoven
    • Contract Manufacturing & Co-Packing
    • Packaging & Food-Contact Materials
    • Regulated Warehousing & Distribution
PricingResourcesCompany
Start free trial Log in Talk to us
Readiness guide · FDA device PCCPs

PCCPs for Devices: Planning Changes Before FDA Authorisation

Define planned changes, tests and labeling in the submission, then prove each release fits the authorised plan.

5 min readFor regulatory, quality and engineering leads at device manufacturers planning iterative design or software changes for the US market
Operator in hair cover, coverall, face mask, safety glasses and blue gloves carrying a tray of bagged moulded components past packed pallets

In Short

  1. 1In-plan changes skip a new 510(k).
  2. 2Out-of-plan: usual route applies.
  3. 3Post-plan versions aren't predicates.
Contents
  1. What the statute allows
  2. What FDA can require in a plan
  3. The Predicate Rule
  4. Worked Example: Plan Record PC-14
  5. Quality System Links
  6. Questions
Contents
  1. What the statute allows
  2. What FDA can require in a plan
  3. The Predicate Rule
  4. Worked Example: Plan Record PC-14
  5. Quality System Links
  6. Questions

Section 515C of the Federal Food, Drug, and Cosmetic Act, codified at 21 U.S.C. 360e-4 and added in December 2022, lets FDA approve or clear a predetermined change control plan (PCCP) as part of a device's marketing submission. Changes made consistent with an authorised plan do not need a new PMA supplement or 510(k). PCCPs are often discussed for machine-learning software, but the statute covers devices generally. FDA's PCCP guidance documents were not opened for this guide. This is orientation, not legal advice. Sources checked 4 Oct 2026.

01

What the statute allows

For approved devices, 21 U.S.C. 360e-4(a) says a supplemental application is not required for a change consistent with an approved PCCP.

Approved vs Cleared Devices under 360e-4

PathwayChange consistent with planCondition
PMA (approved)No supplement neededDevice remains safe and effective
510(k) (cleared)No new notification neededSafe, effective and substantially equivalent
Change outside planUsual route appliesRegulatory team decides
  • PMA (approved)
    Change consistent with plan
    No supplement needed
    Condition
    Device remains safe and effective
  • 510(k) (cleared)
    Change consistent with plan
    No new notification needed
    Condition
    Safe, effective and substantially equivalent
  • Change outside plan
    Change consistent with plan
    Usual route applies
    Condition
    Regulatory team decides
Summary of subsections (a) and (b).
More on what the statute allows

FDA may approve a plan that describes planned changes that would otherwise need a supplement, if the device remains safe and effective without any change. For cleared devices, subsection (b) says a new premarket notification is not required for a change consistent with a cleared plan, provided the device remains safe and effective and would remain substantially equivalent to the predicate. The plan is part of the authorisation, not a later side agreement.

02

What FDA can require in a plan

Under subsections (a)(3) and (b)(3), FDA may require a plan to include labeling required for safe and effective use as the device changes, notification requirements if the device does not function as intended under the plan, and performance requirements for changes made under it. In practice that means each planned change needs defined acceptance criteria, a verification and validation method, and a labeling update approach before submission.

More on what FDA can require in a plan

Changes outside the plan's description follow the usual supplement or 510(k) route; the plan does not widen itself.

03

The Predicate Rule

Subsection (c) limits how changed devices can be used as predicates. When FDA determines substantial equivalence, it shall not compare a device to versions changed under a PCCP as the predicate.

More on the Predicate Rule

Only the version cleared or approved before changes under the plan may be used. For a manufacturer this matters when planning a product family: a later 510(k) cannot rely on the post-PCCP performance of an earlier device as its predicate, so keep a clear record of which version was authorised and which versions resulted from the plan.

04

Worked Example: Plan Record PC-14

Illustrative only. A device software team records plan PC-14 inside its design history: three planned modification types, each with its own test protocol, acceptance criteria and labeling text to update.

Release check against plan PC-14 (illustrative)

0 of 6 in place
Invented example.
More on worked Example: Plan Record PC-14

When release 2.3 is ready, engineering links the verification report, quality confirms the change matches modification type 2 and meets its criteria, and regulatory confirms no step falls outside the authorised description. Only then does the release proceed, with the device master record and labeling version updated. A change that fails the criteria returns to the normal premarket route.

05

Quality System Links

A PCCP does not remove design controls, change control or complaint handling under FDA's quality system requirements; it changes whether a new premarket submission is needed. Treat each implemented change as a controlled design change with its own records, and monitor field performance so the notification requirements in the plan can be met.

More on quality System Links

Decisions about whether a change fits the plan, and whether the device remains safe and effective, are made by qualified people. FDA's final guidance documents on PCCPs, including for AI-enabled software, add recommendations this guide did not review.

How V5 supports this

V5 keeps plan documents, design change records and device history versions linked. Your regulatory and quality people decide whether a change fits the plan.

  • Document controlPlan, protocols and labeling under version control.
  • QMSDesign change records reviewed by people.
  • Device History RecordsVersion history per released configuration.

Questions

What is a predetermined change control plan?

A plan, authorised in a PMA or 510(k), describing planned device changes that can then be made without a new supplement or premarket notification, under 21 U.S.C. 360e-4.

Is a PCCP only for AI software?

No. The statute applies to devices generally. FDA has published guidance focused on AI-enabled software functions, which was not opened for this guide.

Can a device changed under a PCCP be a predicate?

No. Under subsection (c), only the version cleared or approved before changes under the plan may be used as a predicate.

What happens if a change falls outside the plan?

It is not covered by the PCCP, so the usual PMA supplement or 510(k) requirements apply. Your regulatory team decides the route.

Terms
  • PCCP
  • SaMD
Industry
  • Medical Devices
  • Blood & Tissue
  • Pharmaceutical

Practical orientation, not legal advice. Rules differ by jurisdiction; your quality and compliance team decides what applies.

Keep reading

  • Australia Medical Device Regulatory Readiness Guide
  • Brazil medical device regulatory readiness guide (ANVISA)
  • Canada Medical Devices Regulations (SOR/98-282) Readiness
  • China Medical Device Regulatory Readiness Guide 2026
  • eDHR Implementation: A Medical-Device Rollout Guide
  • EU MDR medical device regulatory readiness (European Union)
All guides

Showing the site for the United States. Local spelling, dates and prices are applied automatically; change the language at the top of the page.

V5 Ultimate
Ultimate

Warehouse, quality and manufacturing software for regulated operations.

ProductIndustriesPricingResourcesSecurity & TrustCompanyLegal Centre

© S.G. Systems, LLC

Page version 1.4· Last revised 4 Oct 2026